SenSmart 4100 ==> EC (Electrochemical)
  • Sensor Type:  Electrochemical
  • Detection Range:  Varies with gas type.  (see "Detectable-Gases" tab)
  • Resolution:  0.01 ppm; 0.1 ppm; 1 ppm; (depends on gas type)
  • Response Time:  less than 30 sec. 90%  (for most gases)
  • Recovery Time:  less than 30 sec. 10% (for most gases)
  • Temperature Range:  -40 deg. C to +50 deg. C (see specific data sheet)
  • Humidity Range:  15 - 90% RH, non-condensing (for most gases)
  • Zero Drift:  less than 1% of full-scale typical (for most gases)
  • Power Requirements:  10-30VDC at < 0.75 watts
  • Housing:  Aluminum with epoxy paint standard. Optional #316 stainless steel. Optional Ul94 Poly Black Plastic Enclosure
  • Output:  4-20 mA, 2-Wire
    • Certifications:  CSA Certification for:
      Class I, Div1, Group A, B, C, D , Class I, Zone 1, Group IIC, T4
      Class I, Div 2 Groups A, B, C, D, Class I, Zone 2, Group IIC, T4
    • Sensor Service Life:  >2 years typical under normal conditions depending upon sensor type and conditions
    • Warranty:  5 year limited warranty
      For sensor warranty see sensor specifications sheet.