GD-70 Sample Draw Gas Detector - SPECIFICATIONS

GD-70 Sample Draw Gas Detector - SPECIFICATIONS
by RKI Instruments

MAIN UNIT
Model GD-70D GD-70D-NT GD-70D-ET
Communication 4-20mA DC DC power line communication PoE Method
Detection principle Different type depending upon sensor unit and detectable gas (see table)
Sampling method Sample drawing types (auto-adjustment of flow rate) 0.5 / min +/-10%
Display
  • Large LCD display (white backlight)
  • Gas concentration
  • Flow rate, communication status, Pyrolyzer status, gas detected
  • Error code, content of error
Gas alarms Two alarm levels: 1st alarm – Red 2nd alarm – Red Fault alarm: Yellow
External output 1st, 2nd, and trouble alarms : Relay contact output for each alarm
Self diagnostic function System failure, sensor failure, flow failure, communication error NT / ET / Analog
Data logging Event history, alarm history, calibration history. Alarm trend (180 sec before / after 1st alarm)
Operation temp. & humidity 0 ~ 40ºC, 30 ~ 70% RH (non-condensing)
Operating settings All operational settings are user adjustable through front panel
Power requirements DC 24V+/- 10%, approx 1.5W (Max 4W including sensor unit) Note: Approx. 2.5W (Max 5W) with SGU sensor unit PoE standard arrangement
Dimensions 2.8”W x 4.7”H x 5.9”D (70W x 120H x 150Dmm)
Weight Approx. 0.9kg (2.0lbs), including sensor unit
Mounting Wall-mounting base plate by 2 or 3 screws
Sampling tubing 4 x 6mm PTFE tubing recommended. Tube fittings provided as standard accessories
Bushing Cable type varies depending on communication method (Cable bushing optional)
SENSOR UNIT
Model ESU SGU SSU NCU
Detection Principle Electro-chemical cell Semi-conductor Pyrolysis-particle Catalytic combustion
Gas detected and detection range Refer to list of detectable gases 0-2000ppm H2, CH4, or CH2F2 (R-32) in air and others 0-15ppm TEOS in air 0-100% LEL H2, CH4, and others
Self diagnosis function Sensor trouble, system failure
Date logging function Event history, alarm history, calibration history, Alarm trend (60 sec. before/after 1st alarm)
PYROLYSER UNIT
Model PLU-70
Application NF3 / TEOS gases detected in air
Usage Used by connecting to “GD-70D” (Main unit)
Power Lamp
  • LED (Green color)
  • Normal: Light-on
  • Warming-up: Flashing at every 1 sec interval
  • Trouble: Flashing at every 0.2 sec interval
Self-diagnostic function
  • Pyrolyser unit trouble
  • Fan trouble
  • System trouble
Operating temp. & humidity 0-40° C, 30-70% RH (non-condensing)
Operational settings All operational settings are user adjustable through front panel.
Power requirements DC 24V+/- 10%, approx. 25W (max)
Dimensions 2.8”W x 4.7”H x 5.9”D (70W x 120H x 150Dmm)
Weight Approx. 1.2kg (2.6lbs)
Mounting Wall-mounting base plate by 2 or 3 screws
Sampling 4x6mm PTFE tubing recommended. Tube fittings provided as standard accessories
Bushing 1.25sq 2 core cable for power supply DC24V (Cable bushing optional)