GX-Force Sample Draw 4 Gas Monitor - SPECIFICATIONS

GX-Force Sample Draw 4 Gas Monitor - SPECIFICATIONS
by RKI Instruments

Sampling Method Internal sample pump, flow rate nominal 0.5 LPM, includes hydrophobic filter
Suction Flow Rate Minimum 0.35 L/min (open flow rate)
Gas Alarm Pattern Lamp flashing, continuous modulating buzzer sounding, gas concentration display flashing, vibration
Gas Alarm Reset Operation Self-resetting or latching
Fault Alarm/Self-Diagnosis System, clock, or sensor abnormality; battery voltage drop; calibration failure;
pump abnormality; low flow rate
Fault Alarm Pattern Lamp flashing, intermittent buzzer sounding, detail display
Fault alarm reset operation Self-resetting
Display LCD digital (7-segment + 14-segment + icons) with backlight
Individual Operations Operational status, clock, battery level, peak reading, pump status, calibration notification
Sound Pressure Approx. 90 dB (30 cm)
Data Logger Function Maximum storage capacity: 3,600 items Interval: 5 minutes (adjustable)
Communication Specifications USB, 2.0 (for data logger) Connector: Type-C
Power Source Rechargeable lithium-ion battery, USB Type-C power cord
Continuous Operating Time Approx. 30 hours (25 °C, fully charged, no alarm, no backlight)
Operating Ambient
Temperature/Humidity Range
-20°C to +50 °C (no sudden changes),  0–95 RH (no condensation)
(-4 °F to +122 °F)
Approvals Intrinsically safe construction
IECEx (Ex da ia IIC T4 Ga/Ex ia IIC T4 Ga)
ATEX (II1G Ex da ia IIC T4 Ga/II1G Ex ia IIC T4 Ga)
Certifications

CQPSUS, Class I, Div 1, Groups ABCD

Ex da ia IIC T4 Ga (w/ NCR-6309 sensor)/Ex ia IIC T4 Ga (w/o NCR-6309 sensor)

Class I, Zone 0

AEx da ia IIC T4 Ga (w/ NCR-6309 sensor)/AEx ia IIC T4 GA (w/o NCR-6309 sensor)

Protection Level IP67 equivalent
External Dimensions/Weight Approx. 2.52 (W) × 6.81 (H) × 1.85 (D) (excluding protrusions) / Approx. 9.87 oz.
Warranty Three years material and workmanship

 

Detection target gas

Combustible gas
(CH4 or HC)

Oxygen
(O2)

Carbon monoxide
(CO)

Hydrogen sulfide
(H2S)

Detection principle

Catalytic combustion

Electrochemical type

Detection range

0 – 100 % LEL

0.0 – 40.0 vol %

0 – 2,000 ppm

0.0 – 200.0 ppm

Resolution

1% LEL

0.1 vol %

1 ppm

0.1 ppm

Alarm setpoints

 

(User-defined setting)

1st alarm: 10 % LEL

 

2nd alarm: 25 % LEL

3rd alarm: 50 % LEL

OVER alarm: 100 % LEL

L alarm: 19.5 vol %

 

LL alarm: 18.0 vol %

H alarm: 23.5 vol %

OVER alarm: 40.0 vol %

1st alarm: 25 ppm

 

2nd alarm: 50 ppm

3rd alarm: 1200 ppm

TWA alarm: 25 ppm

STEL alarm: 200 ppm

OVER alarm: 2,000 ppm

1st alarm: 5.0 ppm

 

2nd alarm: 30.0 ppm

3rd alarm: 100.0 ppm

TWA alarm: 1.0 ppm

STEL alarm: 5.0 ppm

OVER alarm: 200.0 ppm

Response time (T90)

CH4: Within 30 seconds,

HC: Within 40 seconds

Within 20 seconds

Within 30 seconds

Within 30 seconds