Aqua TROLL pH/ORP Sensor (P/N 0063470) - OVERVIEW
Aqua TROLL pH/ORP Sensor (P/N 0063470) - OVERVIEW
by In-Situ
The pH/ORP sensor is compatible with the Aqua TROLL 500 and Aqua TROLL 600 multiparameter sondes. It features a single-junction, three-electrode design with refillable reference solution.
This combination pH sensor and ORP sensor has been redesigned to significantly increase sensor stability and data accuracy while minimizing maintenance.
Features:
- Glass electrode pH sensor
- pH measurement to 0.01 pH resolution with accuracy of +/-0.1 pH unit or better
- Long-term stability: The Aqua TROLL pH sensor change is 0.0008 pH/day, or a change of 0.1 pH in 125 days! Learn more »
- ORP probe accuracy of +/-5mV across range of +/-1400mV
- pH electrode consists of pH-sensitive glass with voltage proportional to hydrogen ion concentration
- Uses potentiometric method for pH measurement
- ORP sensor uses a potassium chloride (KCl) solution
Benefits:
- Accurate water quality data: Reference electrode supplies a stable output for measuring ORP and pH in a wide range of applications.
- Minimal maintenance: Increased pH/ORP sensor stability, refillable reference solution, and replaceable reference junction streamline maintenance.
- Simplified data collection: Suitable for remote monitoring via telemetry, wired, or wireless installation methods.